Intermittent Comms - X800

Intermittent Comms - X800

Had a programmer which was reporting USB errors during diagnostics.


usberror


The word from bench techs at BPM:

Have you tried bypassing the internal USB cable with y our external one to see if com is restored reliably using a different USB cable straight to the site?

You may also have an issue with U2 on the site motherboard (LPCBD04143).
Sometimes this small QFP takes a crap and com. will become intermittent as the site heats up ( this happens on older units).
This chip is next to the J10 USB receiver on the MB.


I would also check the termination of the connector at J5 on the power board (LPCBD05056). Sometimes
this can come loose, or back out of the connector.



    • Related Articles

    • X Axis Timeout Error on 4000 Series Windows Autohandler

      The X-axis is the long, left/right axis. Also called the upper axis. (Note: For old DOS systems, this would be the "Y" axis)  The time-out error message is received because the X motor drive system did not get to the encoder position within a certain ...
    • X Axis Timeout Error on 4000 Series Windows Autohandler

      The X-axis is the long, left/right axis. Also called the upper axis. (Note: For old DOS systems, this would be the "Y" axis)  The time-out error message is received because the X motor drive system did not get to the encoder position within a certain ...
    • Internal Runtime error

      Error appeared on a BP3800MK2 with Galil 4040 and OS Win 7 32 bit when BPwin has been updated from V5.55 to V7.0.x  "Internal Runtime Error" message are generic, and do not mean anything other than BPWin and Windows have a conflict.  ...
    • BP- 3000 Series Autohandler Maintenance

      Weekly – 10 minutes Dust off programmer sites and socket modules with clean, dry compressed air. Inspect Z-rod bellows for proper movement (compression and extension). Check nozzle runout. Replace bellows if run-out is greater than 8 mils. Remove ...
    • How to change the IP address on FR Cube

      In this mini guide, I go through how you can change the IP address on a FlashRunner Cube, to enable communication through the Ethernet port. Firstly, you have to know the IP address of the programmer you wish to access. If you do not know the IP ...